| | IRA-VISIR | IRA-MIR |
| Full possible input wavelength range* | 450-2000 nm | 2-11 um |
| Subranges* | VIS: 450-700 nm NIR1: 700-1300 nm NIR2: 1300-2000 nm | MIR1: 2-5 um MIR2: 5-11 um |
| Input pulse duration range | 50 fs - 250 ps |
| Required input pulse energy** | >1 nJ at 50 fs - 1 ps (w. thin NL crystals) >3 nJ at 1-250 ps (w. thick NL crystals) | >5 nJ at 50 fs - 1 ps (w. thin NL crystals) >100 nJ at 1-250 ps (w. thick NL crystals) |
| Maximum input average power | 1 W |
| Input repetition rate | 10 Hz - 100 MHz |
| Input polarization | linear, horizontal |
| Delay line temporal resolution | 8.3 fs |
| Full scan range | 850 ps |
| Required equipment | PC with USB; Windows acquisition and analysis software included |
| Power supply | 220/110 V; 50/60 Hz ±10% |
| Dimensions | optical unit: 450x250x210 mm control unit: 250x180x90 mm |
| * - each subrange is covered by an exchangeable optics set (NL crystals, beamsplitters, filters, photodetectors). A set for one of the subranges of the customer's choice is supplied with the unit, additional sets are supplied upon request; the final set of optics and detectors depends on the specifications of the sources to be measured and is discussed with our sales manager upon offering; ** - typical values, depends on input pulse duration and wavelength. |